Adwa AD32 Bruksanvisning
Sidor 2 Engels Mätutrustning
Läs bruksanvisningen för Adwa AD32 gratis online eller ladda ner PDF:en. 23 personer har tyckt att guiden var hjälpsam. Fråga om modellen? Ställ en fråga

www.adwainstruments.com
USER MANUAL
AD31 & AD32
Waterproof EC/TDS Testers
Dear Customer,
Thank you for choosing an Adwa product.
Please read carefully this manual before
starting operations.
This instrument is in compliance with the
EMC directive 2004/108/EC and its stan-
dards, and Low Voltage Directive 2006/95/
EC and its standards for electrical equip-
ments.
For additional technical information, please
e-mail us at [email protected].
INTRODUCTION
AD31 and AD32 are waterproof EC, TDS and
temperature testers. The housing has been
completely sealed against humidity.
All EC and TDS readings are automatically
temperature compensated (ATC), and
temperature values can be displayed in °C
or °F units.
The EC/TDS conversion factor (CONV) can
be selected by the user, as well as the coefficient
β (BETA) for temperature compensation. The
instruments can be calibrated at one point.
Measurements are highly accurate with a
unique stability indicator right on the LCD.
The models are also provided with a low
battery symbol which warns the user when
the batteries need to be replaced.
The AD32P probe supplied with the meters,
is interchangeable and can be easily replaced
by the user.
The encapsulated temperature sensor allows
fast and accurate temperature measurement
and compensation.
Each meter is supplied complete with:
•AD32P EC probe
•4 x 1.5V batteries, button type
•User manual
FRONT PANEL & DISPLAY
1.Dual line LCD
2.ON/OFF/ MODE button
3.EC probe & temperature sensor
4.Probe body
5.Battery compartment (inside)
6.SET/HOLD button
7.Clip holder
1.Percentage indicator for battery life
2.Stability indicator (hourglass symbol)
3.Low battery warning indicator
4.Calibrated meter indication
5.ATC (Automatic Temperature Compensation)
indicator
6.Secondary LCD level
7.Primary LCD level
8.Measurement unit for primary level
(ppm & μS for AD31, ppt & mS for AD32)
ISTAD31 09/14
TECHNICAL DATA
Range0.0 to 60.0 °C / 32.0 to 140.0 °F
0 to 3999 μS/cm / 0 to 2000 ppm (AD31)
0.00 to 20.00 mS/cm / 0.00 to 10.00 ppt (AD32)
Resolution0.1 °C / 0.1 °F
1 μS/cm / 1 ppm (AD31)
0.01 mS/cm / 1 ppt (AD32)
Accuracy±0.5 °C / ±1 °F
(@25 °C/77 °F)±2% f.s. (EC/TDS)
CalibrationAutomatic, 1 point
1413 μS/cm, 1382 ppm or 1500 ppm (AD31)
12.88 mS/cm, 6.44 ppt or 9.02 ppt (AD32)
TDS Factor
User-selectable from 0.45 to 1.00
Temperature Compensation
Automatic, from 0 to 60 °C
β adjustable from 0.0 to 2.4%/°C
ProbeAD32P (included)
Battery Type4 x 1.5V button type
Battery LifeApprox. 100 hours of use
Auto-offAfter 8 minutes of non-use
Environment
-5 to 50 °C (23 to122 °F); RH 100%
Dimensions175.5 x 39 x 23 mm
Weight100 g
AD32PSpare EC probe for AD31 &
AD32
AD70031P1413 μS/cm EC standard solution
20 ml sachet (25 pcs.)
AD70030P12.88 mS/cm EC standard solution
20 ml sachet (25 pcs.)
AD70032P1382 ppm solution TDS standard
20 ml sachet (25 pcs.)
AD70442P1500 ppm solution TDS standard
20 ml sachet (25 pcs.)
PROBE & SOLUTIONS
Produktspecifikationer
| Varumärke: | Adwa |
| Kategori: | Mätutrustning |
| Modell: | AD32 |
Behöver du hjälp?
Om du behöver hjälp med Adwa AD32 ställ en fråga nedan och andra användare kommer att svara dig
Mätutrustning Adwa Manualer
Adwa AD330 Bruksanvisning Adwa AD3000 Bruksanvisning Adwa ECO407 Bruksanvisning Adwa AD14 Bruksanvisning Adwa AD130 Bruksanvisning Adwa AD131 Bruksanvisning Adwa AD203 Bruksanvisning Adwa AD201 Bruksanvisning
Alla Adwa Mätutrustning manualer
Populära i Mätutrustning
Mätutrustning Manualer
Nyaste Mätutrustning Manualer
Kern ORM 2SU Bruksanvisning Kern KFP 60V30LM Bruksanvisning Kern ORA 4AB Bruksanvisning Grundfos Hydro Multi-S Bruksanvisning Sauter FH 500 Bruksanvisning Sauter FH 100 Bruksanvisning Sauter TVS 20KN100 Bruksanvisning Sauter CJ P4PG Bruksanvisning Sauter DA 5-3 Bruksanvisning Sauter SD 50N100 Bruksanvisning