Läs gratis den bruksanvisning för Fluke 700P00 (4 sidor) i kategorin Inte kategoriserad. Guiden har ansetts hjälpsam av 7 personer och har ett genomsnittsbetyg på 4.9 stjärnor baserat på 4 recensioner. Har du en fråga om Fluke 700P00 eller vill du ställa frågor till andra användare av produkten? Ställ en fråga

Sida 1/4
®
PN 942735 July 1994 Rev.12, 11/01
©1994-2001 Fluke Corporation. All rights reserved. Printed in USA
All product names are trademarks of their respective companies
700 Series Pressure Modules
Instruction Sheet
15 PSID/G
PRESSURE MODULE
700
P
04
1 bar100 kPa
RANGE
Low
High
Differential
100 PSIG
PRESSURE MODULE
700
P
06
7 bar700 kPa
RANGE
Gage
yz1f.eps
Figure 1.
Introduction
The Fluke 700 Series Pressure Modules allow you to measure
pressure with the Fluke 700 Series Process Calibrators. Read this
sheet before you use the pressure module. This sheet contains
specifications and information about how to avoid damaging the
pressure module. See the Calibrator Users Manual for operating
instructions.
The pressure module measures pressure using an internal
microprocessor. It receives operating power from and sends digital
information to the 700 Series calibrator.
Gage pressure modules have one pressure fitting and measure
pressure with respect to atmospheric pressure. Differential pressure
modules have two pressure fittings and measure the difference
between the applied pressure on the high fitting versus the low fitting.
Figure 1 shows the two types. A differential pressure module functions
like a gage module when the low fitting is open.
Compatibility of Pressure Modules with Fluke
Process Calibrators
Fluke-701, 702, 741, 743, and 744 Documenting Process
Calibrators: The Fluke-700P00 and P27 Pressure Modules are
completely compatible with these calibrators.
Fluke-716, 717, and 718 Pressure Calibrators: The Fluke-700P00
and P27 Pressure Modules will display pressure readings correctly on
716 Pressure Calibrators with firmware V1.2 or higher, or 717 and 718
Pressure Calibrators with firmware V1.3 or higher. To view the software
revision, power-on the calibrator while holding down the MAX key.
Read the REV number in the display. Contact your Fluke Service
Center for a free update of calibrators with lower revision numbers.
Reference PCN 4647.