Läs gratis den bruksanvisning för IFM PC3542 (5 sidor) i kategorin Inte kategoriserad. Guiden har ansetts hjälpsam av 7 personer och har ett genomsnittsbetyg på 4.3 stjärnor baserat på 4 recensioner. Har du en fråga om IFM PC3542 eller vill du ställa frågor till andra användare av produkten? Ställ en fråga
Sida 1/5

Installation Instructions
Electronic pressure sensor
for industrial applications
PC3541
PC3542
704775 / 00 04 / 2010
UK

2
1 Safety instructions
Please read this document prior to installing the unit. Ensure that the product is •
suitable for your application without any restrictions.
If the operating instructions or the technical data are not adhered to, personal •
injury and/or damage to property may occur.
Please check for all applications that the product materials (see Technical data) •
are compatible with the media to be measured.
For the scope of validity cULus:
The device shall be supplied from an isolating transformer having a secondary
Listed fuse rated either
a) max 5 amps for voltages 0~20 Vrms (0~28.3 Vp) or
b) 100/Vp for voltages of 20~30 Vrms (28.3~42.4 Vp).
Contents
1 Safety instructions ................................................................................................2
2 Function and features ...........................................................................................3
2.1 Applications ....................................................................................................3
3 Installation ............................................................................................................3
4 Electrical connection.............................................................................................4
5 Scale drawing .......................................................................................................4
6 Technical data .......................................................................................................5

3
UK
2 Function and features
The pressure sensor detects the system pressure and converts it into an analog
output signal.
4 ... 20 mA•
2.1 Applications
Type of pressure: relative pressure •
Order no. Measuring range Permissible
overload pressure
Bursting
pressure
bar PSI bar PSI bar PSI
PC3541 0...250 0...3 625 400 5 800 1 000 14 504
PC3542 0...100 0...1 450 200 2 900 1 000 14 504
MPa = bar ÷ 10 / kPa = bar × 100
Static and dynamic overpressures exceeding the indicated overload pres-
sure are to be avoided by taking appropriate measures.
The indicated bursting pressure must not be exceeded. Even if the burst-
ing pressure is exceeded only for a short time, the unit can be destroyed.
NOTE: Risk of injury!
3 Installation
Before mounting and removing the sensor, make sure that no pressure is
applied to the system.
Insert the unit in a G¼ process connection. ►
Tighten firmly. T ightening torque: max. 25 Nm. ►
Produktspecifikationer
Varumärke: | IFM |
Kategori: | Inte kategoriserad |
Modell: | PC3542 |
Behöver du hjälp?
Om du behöver hjälp med IFM PC3542 ställ en fråga nedan och andra användare kommer att svara dig
Inte kategoriserad IFM Manualer
8 April 2025
8 April 2025
8 April 2025
2 April 2025
2 April 2025
1 April 2025
1 April 2025
1 April 2025
31 Mars 2025
12 Mars 2025
Inte kategoriserad Manualer
- Inglesina
- Prime3
- K&K Sound
- Nautilus
- MBM
- Grothe
- Lowrance
- Edbak
- Campomatic
- SPT
- Canicom
- Wilson
- Lascar Electronics
- Samlex
- Brondi
Nyaste Inte kategoriserad Manualer
9 April 2025
9 April 2025
9 April 2025
9 April 2025
9 April 2025
9 April 2025
9 April 2025
9 April 2025
9 April 2025
9 April 2025