    Läs gratis den bruksanvisning för Omron HEM-7113 (1 sidor) i kategorin Blodtrycksmätare. Guiden har ansetts hjälpsam av 9 personer och har ett genomsnittsbetyg på 4.4 stjärnor baserat på 5 recensioner. Har du en fråga om Omron HEM-7113 eller vill du ställa frågor till andra användare av produkten? Ställ en fråga
                        Sida 1/1
                    
                    
                    
HEM-7113-AP_C_M03_120219.pdf
Instruction Manual
Introduction
Important Safety Information
1.   Overview
Model  HEM-7113
The unit also stores up to 14 measurements in memory.
Main unit: 
A. Display
C. Memory button
D. Battery compartment
E. Air jack
A
B C
D
E
5337886-7A
Thank you for purchasing the OMRON HEM-7113 Digital Automatic Blood 
Pressure Monitor.
The OMRON HEM-7113 is a fully automatic blood pressure monitor, operating 
on the oscillometric principle. It measures your blood pressure and pulse rate 
simply and quickly.
Please read this instruction manual thoroughly before using the unit. For 
specific information about your own blood pressure, CONSULT YOUR 
DOCTOR.
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia or 
arteriosclerosis.
Please read this section carefully before using the unit.
     Warning:
• Indicates a potentially hazardous situation which, if not avoided, could result in 
death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and 
self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a 
doctor before using the unit, as cuff inflation can cause internal bleeding.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with plenty of clean 
water. Consult a doctor immediately.
     Caution:
• Indicates a potentially hazardous situation which, if not avoided, may result in 
minor or moderate injury to the user or patient or damage to the equipment or 
other property.
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express 
their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or arm cuff.
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result in 
incorrect measurement results.
• Make sure that the air tube is not wrapped around other parts of your body when 
taking measurements. This could result in injury when the air pressure in the air 
tube is increased.
• Do not leave the cuff wrapped on the arm if taking measurements during the 
night. This could result in injury.
• Do not inflate the arm cuff over 299 mmHg.
• Do not use a mobile phone or other devices that emit electromagnetic fields, 
near the unit. This may result in incorrect operation of the unit.
• Do not operate the unit in a moving vehicle (car, airplane).
• To inflate the cuff manually, refer to Chapter 3.3. If the cuff is over inflated, it can 
cause internal bleeding.
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse with plenty of 
clean water.
• Use only four“ AAA” alkaline batteries with this unit. Do not use other types of 
batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace all four batteries at 
the same time.
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together. General Precautions
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not take measurements after bathing, drinking alcohol, smoking, exercising 
or eating.
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• When removing the air tube, pull on the air plug at the connection with the main 
unit not the tube itself.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not wash the arm cuff or immerse it in water.
• Read and follow the“ Important information regarding Electro Magnetic 
Compatibility (EMC)” in the Technical Data Section.
• Read and follow the“ Correct Disposal of This Product” in the Technical Data 
Section when disposing of the device and any used accessories or optional 
parts.
B. START/STOP button.
Automatic 
Blood Pressure Monitor
Be sure to wrap the arm cuff correctly so that you get accurate results.
Measurements can be taken in light clothing. However, please remove 
thick clothes, such as sweaters, before taking a reading.
2.1   Installing/Replacing the Batteries
2.   Preparation
3.   Using the Unit
2.2   Battery Life & Replacement
3.1   Applying the Arm Cuff
1. Turn the main unit upside down.
2. Slide the battery cover in the direction 
of the arrow while pressing the ribbed 
part of the cover.
3. Install or replace four “AAA” size 
batteries so that the + (positive) 
and - (negative) polarities match 
the polarities indicated on the battery 
compartment.
4. Put the battery cover back in place.
Slide the battery cover as indicated, 
until it clicks into place. 
- When the battery low symbol (     ) starts to blink, you will still be able to use 
  the unit for a short while. You should replace the batteries with new ones 
  ahead of time.
- When the symbol (     ) remains lit, the batteries are exhausted. You should 
  replace the batteries with new ones at once. Turn the unit off before replacing 
  the batteries.
Four new “AAA” alkaline batteries will last for approximately 300 measurements, 
when used to take two measurements a day.
Since the supplied batteries are for monitoring use only, they may have 
a shorter life and not last for 300 measurements.
• Remove the batteries if the unit will not be used for three months or more.
• Dispose of batteries according to applicable local regulations.
If the battery low symbol (     ) appears on the display, 
replace all  four batteries at the same time.
1.
1. To delete stored readings, first press the Memory (M) 
button.
Then while holding it down, press the START/STOP 
button simultaneously for about 2-3 seconds. All readings 
will then be deleted.
2. Press the START/STOP button to turn the unit off.
If you forget to turn the unit off, it will automatically shut
itself off after five minutes.
To Delete All the Values Stored in Memory
You cannot delete individual stored readings, all the
readings in the unit will be deleted.
21
4.   Handling Errors and Problems
4.1   Error Messages
Error Display Cause Remedy
Cuff is under inflated.Carefully read and repeat the 
steps listed in section 3.3.
Movement during 
measurement.
Repeat measurement. Remain 
still and do not talk during 
measurement.
Air plug disconnected. Insert the air plug securely.
Refer to section 3.1.
Arm cuff not applied 
correctly.
Apply the arm cuff correctly. 
Refer to section 3.1.
Clothing is interfering 
with the arm cuff.
Remove any clothing 
interfering with the arm cuff.
Air is leaking from the 
arm cuff. Replace cuff with new one.
Arm cuff has been 
inflated above 
299 mmHg.
Do not inflate the arm cuff 
above 299 mmHg.
Refer to section 3.3.
Intended Use
This product is designed to measure the blood pressure and pulse rate of 
people within the range of the designated arm cuff, following the instructions in 
this instruction manual.
It is mainly designed for general household use. Please read the Important 
Safety Information in this instruction manual before using the unit.
1 - 2 cm
2.
Note: 
You can take a measurement on either your left or right arm. The blood 
pressure can differ between the right arm and the left arm and therefore 
also the measured blood pressure values can be different. OMRON 
recommends to always use the same arm for measurement. If the 
values between the two arms differ substantially, please check with your 
physician which arm to use for your measurement.
Insert the air plug into the air jack 
on the left side of the main unit.
Apply the arm cuff to your upper arm. The air tube should be 
centred on the inside of your arm and point down the inside of the 
arm, so that the air tube runs down the inside of your forearm and 
is in line with your middle finger. 
3.3   Instructions for Special Conditions
5. Press the START/STOP button to turn the unit off.
1. Press the START/STOP button 
to turn the unit on.
Measurement starts.
Note: If you forget to turn the unit off, it will automatically shut itself off after 
five minutes.
If your systolic pressure is known to be more than 220 mmHg, press 
and hold the START/STOP button until the arm cuff inflates 30 to 
40 mmHg higher than your suspected systolic pressure.
Arm cuff: 
F.  Arm cuff
(Medium cuff:
  arm circumference 22-32 cm)
G. Air tube
Note: 
GF
H
Package contents: 
Display: 
J. Systolic blood pressure
K. Diastolic blood pressure
L. Memory symbol
N. Heartbeat symbol
  1. Flashes during measurement
O. Pulse display
P. Deflation symbol
Q. Battery low symbol
I. 
•Instruction manual
•Blood pressure monitor unit
I
J
LN
P
K
O
Q
H. Air plug
M. Irregular Heartbeat symbol
M
If air starts to leak from the arm cuff, 
please replace with a new one.
Refer to “6. Optional Parts”.
The optional arm cuff does not 
include the air plug. Do not throw 
the air plug away in case 
purchasing the new one.
Four “AAA” alkaline (LR03) 
batteries
 2. If flashing after measurement 
complete indicates blood pressure 
out of recommended range
Note: If your systolic or diastolic 
pressure is outside the 
standard range 
(above 135/85 mmHg), 
the Heartbeat symbol (    ) 
will blink. Refer to 
“3.2 Taking a Reading”.
Important:
•
4. Remove the arm cuff.
Your blood pressure monitor includes an 
irregular heartbeat feature. Irregular heart-
beats can influence the results of the 
measurement. The irregular heartbeat 
algorithm automatically determines if the 
measurement is usable or needs to be 
repeated. If the measurement results are affected by 
irregular heartbeats but the result is valid, the result is 
shown together with the irregular heartbeat symbol (     ).
If the irregular heartbeats cause the measurement to be 
invalid, no result is shown. If the Irregular heartbeat 
symbol (     ) is shown after you have taken a measure-
ment, repeat the measurement. If the Irregular heartbeat 
symbol (     ) is shown frequently, please make your doctor 
aware of it.
NormalHeartbeat
Blood
pressure
Blood
pressure
Pulse
Pulse
     Warning:
What is Arrhythmia?
An irregular heartbeat is a heartbeat 
rhythm that varies by more than 25% 
from the average heartbeat rhythm 
detected while the unit is measuring 
the systolic and diastolic blood 
pressure. If such an irregular rhythm 
is detected more than twice during 
measurement, the irregular heartbeat 
symbol (     ) appears on the symbol 
when the measurement results are 
displayed.
A heartbeat is stimulated by electrical signals that cause the heart to contract.
Arrhythmia is a condition where the heartbeat rhythm is abnormal due to flaws 
in the bio-electrical system that drives the heartbeat. Typical symptoms are 
skipped heartbeats, premature contraction, an abnormally rapid (tachycardia) 
or slow (bradycardia) pulse. This can be caused by heart disease, aging, 
physical predisposition, stress, lack of sleep, fatigue etc. Arrhythmia can only 
be diagnosed by a doctor through a special examination. Whether the 
appearance of the irregular heartbeat symbol (     ) in the results indicates 
arrhythmia or not can only be determined by an examination and diagnosis by 
your doctor.
What is Irregular Heartbeat?
If the irregular heartbeat symbol (     ) is shown frequently, please make 
your doctor aware of it. Conducting self-diagnosis and treatment based on 
measurement results is dangerous. Be sure to follow the instructions of 
your doctor.
Irregular 
Short
Heartbeat 
Long
Note: The measurement values 
continue to be stored in 
memory even after the 
batteries are replaced.
Arm cuff pressure
does not rise. 
Make sure that the air 
jack is connected
securely. 
Refer to section 3.1.
Air is leaking from the
arm cuff. 
Replace the arm cuff 
with a new one. 
Arm cuff deflates too
soon.  The arm cuff is loose.
Apply the cuff correctly 
so that it is firmly 
wrapped around the 
arm. 
Refer to section 3.1.
Cannot measure or
readings are too low
or too high. 
Inflate the cuff so that it 
is 30 to 40 mmHg 
above your previous 
measurement result.
Refer to section 3.3.
The unit loses power 
during measurement.
The batteries are 
empty.
Replace the batteries 
with new ones.
Nothing happens 
when you press the 
buttons.
The batteries are 
empty.
Replace the batteries 
with new ones.
The batteries have 
been inserted 
incorrectly.
Insert the batteries with 
the correct (+/ -) 
polarity.
4.2   Troubleshooting
Problem CauseRemedy
The reading is
extremely low (or 
high). 
Arm cuff not applied 
correctly.
Apply the arm cuff 
correctly. 
Refer to section 3.1.
Movement or talking 
during measurement.
Remain still and do not 
talk during 
measurement.
Clothing is interfering 
with the arm cuff.
Remove any clothing 
interfering with the arm 
cuff.
Other problems.
Press the START/STOP button and repeat 
measurement.
If the problem continues, try replacing the 
batteries with new ones.
If this still does not solve the problem, contact your
local authorized OMRON distributor or dealer.
Note: If the memory is full, the monitor will delete the oldest readings.
This unit has a memory capable of storing 14 sets of readings.
Press the Memory (M) button.
The result from the most recent measurement is displayed.
Note: If there are no measurements results
        stored in memory, the screen to the right 
        is displayed.
If your systolic or diastolic pressure is
outside the standard range, the heartbeat
symbol will blink when the measurement 
result is displayed. 
Refer to “3.2 Taking a Reading”.
Press the Memory (M) button repeatedly to 
cycle through the previous measurement 
results.
The Memory number appears for a second 
before the pulse rate is displayed. 
The newest set is numbered “1”.
Press the START/STOP button to turn the unit off.
If you forget to turn the unit off, it will automatically shut
itself off after five minutes.
3.4   Using the Memory Function
Important:
1.
2.
3.
At the end of the measurement, if an irregular heartbeat 
is detected, the irregular heartbeat symbol will be 
displayed with the result.
1 to 2 cm
3.2   Taking a Reading
Taking a Reading on the Right Arm
3.
Sit on a chair with your feet flat on 
the floor and place your arm on a 
table so that the arm cuff will be 
at the same level as your heart.
Keep still and do not talk during measurement.
2. Press the START/STOP button to turn on the unit.
1.
3.
Check the measurement results.
The unit automatically stores blood 
pressure and pulse rate into its memory.
Refer to “3.4 Using the Memory 
Function”.
Notes:
• Self-diagnosis of measured results and treatment are dangerous. 
  Please follow the instructions of your doctor.
• Wait 2-3 minutes before taking another blood pressure reading. 
  Waiting between readings allows the arteries to return to the condition prior 
  to taking the blood pressure measurement.
Diastolic blood 
pressure
Systolic blood 
pressure
Pulse display
When the cuff is positioned correctly, close the fabric fastener 
firmly.
Apply the cuff so that the air tube is 
at the side of your elbow.
Notes:
• Be careful not to rest your arm on 
the air tube, or otherwise restrict the 
flow of air to the cuff.
• The cuff should be 1 to 2 cm above 
the elbow.
Note: To cancel a measurement, press the START/STOP button to turn off 
the unit and to release the air in the arm cuff.
Error Display Cause Remedy
Battery power is low.
Replace all four “AAA” 
batteries with new ones.
Refer to section 2.1.
Device error.
Blinks or appears
continuously 
Note: The irregular heartbeat symbol (     ) may also be displayed with error messages.
Note: You cannot inflate the cuff above 299 mmHg. (An error will be displayed 
          if you try to inflate the cuff above 299 mmHg.)
Note: Do not apply more pressure than necessary.
2. When the cuff starts to inflate, press the START/STOP button 
and keep it pressed until the pressure is 30 to 40 mmHg 
higher than your suspected systolic pressure.
3. Release the START/STOP button when the cuff has been 
inflated to the desired pressure.
The cuff starts to deflate and measurement starts.
4. The rest of the procedure is the same as for normal 
measurement. Refer to “3.2 Taking a Reading”, steps 3 to 5.
The air plug is not 
securely inserted into 
the main unit.
The arm cuff has 
not been inflated 
sufficiently.
5.   Maintenance and Storage
Maintenance
To protect your unit from damage, please observe the following:
7.   Technical Data
6.   Optional Parts
• The unit should be cleaned with a soft, dry cloth.
• 
Use a soft, moistened cloth and soap to clean the arm cuff.
Calibration and Service
• The accuracy of this blood pressure monitor has been
carefully tested and is designed for a long service life.
Note: Do not bend the air tube excessively.
Unplug the air tube from the air jack.
Gently fold the air tube
into the arm cuff.
Storage
Do not store the unit in the following situations:
• If the unit is wet.
• Locations exposed to extreme temperatures, humidity, direct sunlight, 
dust or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will be at an angle.
Medium arm cuff
Arm circumference 
22 - 32 cm
Large arm cuff
Arm circumference 
32 - 42 cm
• Do not subject the main unit and cuff to extreme 
temperatures, humidity, moisture or direct sunlight.
• Do not fold the cuff or tubing tightly.
• Do not disassemble the unit.
• Do not subject the unit to strong shocks or vibrations 
(for example, dropping the unit on the floor).
• Do not use volatile liquids to clean the main unit. 
• Do not wash the arm cuff or immerse it in water.
• Do not use petrol, thinners or similar solvents to clean 
the arm cuff.
• Do not carry out repairs of any kind by yourself.
  If a defect occurs, consult your local authorized 
OMRON distributor or dealer.
• 
It is generally recommended to have the unit inspected every 
two years to ensure correct functioning and accuracy. Please 
consult your local authorized OMRON distributor or dealer.
Automatic Blood Pressure Monitor
HEM-7113 (HEM-7113-AP)
LCD Digital Display
Oscillometric method
Pressure: 0 mmHg to 299 mmHg
Pulse: 40 to 180/min.
14 Measurements
Pressure: ±3 mmHg
Pulse: ±5% of display reading
Automatic by electric pump
Automatic pressure release valve
4 "AAA" batteries 1.5V
Capacity of new alkaline batteries is approximately
300 measurements
Notes: 
• Subject to technical modification without prior notice.
• 
Disposal of this product and used batteries should be carried out in accordance
  with the national regulations for the disposal of electronic products.
Operating
temperature/ Humidity
Storage temperature/
Humidity/ Air Pressure
Console Weight
Cuff Weight
Outer Dimensions
Cuff Dimensions
Package Content
10°C to 40°C/ 30 to 85% RH
-20°C to 60°C/ 10 to 95% RH/ 700-1060 hPa
Approximately 245 g without batteries
Approximately 120 g
Approximately 
104 (w) mm × 64 (h) mm × 128.5 (I) mm
Approximately 146 mm × 446 mm
(Medium cuff: arm circumference 22 to 32 cm)
Main unit, cuff, Instruction manual, Battery set
Product Description
Model
Display
Measurement Method
Measurement Range
Memory
Accuracy 
Inflation
Deflation
Power Source
Battery life
Applied Part
Protection Against
Electric Shock
= Type B
Internally powered ME equipment
Asia Pacific HQ
Production Facility
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto,
617-0002 JAPAN
OMRON HEALTHCARE SINGAPORE PTE LTD.
OMRON HEALTHCARE MANUFACTURING
VIETNAM CO., LTD.
Binh Duong Province, VIETNAM
438A Alexandra Road, #05-05/08, Alexandra Technopark,
Singapore 119967 
www.omron-healthcare.com.sg
This OMRON product is manufactured under the strict quality system of OMRON 
HEALTHCARE Co., Ltd., Japan. The Core component for OMRON blood pressure 
monitors, which is the Pressure Sensor, is produced in Japan for assembly.
Contact your local 
authorized OMRON 
distributor or dealer
Produktspecifikationer
| Varumärke: | Omron | 
| Kategori: | Blodtrycksmätare | 
| Modell: | HEM-7113 | 
Behöver du hjälp?
Om du behöver hjälp med Omron HEM-7113 ställ en fråga nedan och andra användare kommer att svara dig
Blodtrycksmätare Omron Manualer
                        
                         7 April 2025
                        
                    
                                                            
                        
                         6 April 2025
                        
                    
                                                            
                        
                         6 April 2025
                        
                    
                                                            
                        
                         5 April 2025
                        
                    
                                                            
                        
                         5 April 2025
                        
                    
                                                            
                        
                         4 April 2025
                        
                    
                                                            
                        
                         3 April 2025
                        
                    
                                                            
                        
                         2 April 2025
                        
                    
                                                            
                        
                         31 Mars 2025
                        
                    
                                                            
                        
                         31 Mars 2025
                        
                    
                                                            Blodtrycksmätare Manualer
- Kinetik
- Orava
- Inovalley
- Uebe
- A&D Medical
- Archos
- Perel
- Melissa
- Inventum
- Elta
- A&D
- Thomson
- Duronic
- Microlife
- Emporia
Nyaste Blodtrycksmätare Manualer
                        
                         4 April 2025
                        
                    
                                                            
                        
                         3 April 2025
                        
                    
                                                            
                        
                         31 Mars 2025
                        
                    
                                                            
                        
                         30 Mars 2025
                        
                    
                                                            
                        
                         30 Mars 2025
                        
                    
                                                            
                        
                         28 Mars 2025
                        
                    
                                                            
                        
                         28 Mars 2025
                        
                    
                                                            
                        
                         17 Mars 2025
                        
                    
                                                            
                        
                         14 Mars 2025
                        
                    
                                                            
                        
                         12 Mars 2025