Läs gratis den bruksanvisning för PCE Instruments PCE-MO 1500 (2 sidor) i kategorin Mätutrustning. Guiden har ansetts hjälpsam av 23 personer och har ett genomsnittsbetyg på 4.9 stjärnor baserat på 9 recensioner. Har du en fråga om PCE Instruments PCE-MO 1500 eller vill du ställa frågor till andra användare av produkten? Ställ en fråga
Sida 1/2

User Manual
Megohmmeter
PCE-MO 1500
SafetyRequirements&SafetyWarnings:
Thisseriesofinsulationresistancetesterisdesignedandproducedaccordingtothesafety
standardIEC1010.StrictlyfollowthesafetystandardsofdoubleinsulationCATIII1000V
CATIVandpollutiondegree2.Toprotectyoursafety,pleasereadthemanualbeforeusing
it.
Precautions:
SafetySymbolDescription:
Caution!Refertotheuser’smanualbeforeusingthismeter.DoubleInsulation
Warning!DangerousVoltage(Risk
ofelectricshock).
Danger:Toavoidseriousorfataldamagethatmaybecausedbycertainstatesandwrong
operations.
Warning:indicatesavoidingtheriskofelectricshock.
Caution:indicatesavoidingdamagetotheinstrumentandmakingaccuratemeasurements.
·Donottestinflammablelocations,sparksmaycauseanexplosion.
·Donotoperatetheinstrumentiftheinstrumentsurfaceiswetoriftheoperator'shands
arewet.
·Whentestingvoltage,thereisapossibilityofpersonalinjuryduetoaccidental
short-circuitbetweenmetalpartsandtestleads.
·Donotexceedthemaximumallowablerangeofthemeasuringrangeduringmeasure-
ment.
·Donotopenthebatterycoverduringmeasurement.
·Whenperforminginsulationmeasurement,donottouchthelineundertest.
Warning
·Iftheinstrumentisabnormal,pleasestopusingit.Forexample:Brokeninstrumentor
exposedmetalparts,pleasedonotuseitinthissituation.
·Whenthetestleadisinsertedintotheinstrumentinterfaceandconnectedtothecircuit
fortesting,donotpressthefunctionbutton,butexitthemeasurementfirst,andthen
changethetestitem.·
Donotinstallsubstitutepartsormakeanyunauthorizedchangestotheinstrument.
·Donotreplacethebatterywhentheinstrumentiswet.
·Makesurealltestleadsaresecurelyconnectedtothemeter'stestsockets.
·Makesuretheinstrumentisturnedoffbeforeopeningthebatterycover.
Caution:
·Beforemeasuring,confirmthatthefunctionbuttonisselectedtotheappropriate
position.
·Afteruse,pressandholdthe“OFF”buttontoturnofftheunit.Ifyoudonotuseit
foralongtime,pleasetakeoutthebatteryandstorageit.
·Donotstorageforalongtimeinhightemperature,humidity,placeswherecondensation
mayoccur,orindirectsunlight.
·Pleaseuseadampclothordetergenttocleantheinstrumentcase,donotuseabrasives
orsolventstowipethe
instrumentcaseandaccessories.
·Whentheinstrumentiswet,pleasedryitbeforeuseorstorage.
General:
Insulationresistancetester(orhighvoltageinsulationmeter)isaspecialinstrumentfor
measuringinsulationresistance.Itisanidealtestinstrumentformeasuringinsulation
resistanceoflargetransformers,transformers,generators,high-voltagemotors,power
capacitors,powercables,arresters,etc.
·Automaticreleasevoltagefunction.
·Suit
ableformeasuringinsulationresistanceofvariouselectricalequipmentand
insulatingmaterialssuchastransformers,motors,cables,switches,electricalappliances,etc.
Itissuitableforthemaintenance,repair,testandverificationofvariouselectrical
equipment.
·Ratedoutputvoltagevariable
Theratedoutputvoltagecanbeswitchedbythefunctionbuttons"▲""▼":
1000Vswitchesbetween50V,100V,250V,500V,1000V
2500Vswitchesbetween250V,500V,1000V,2000V,2500V.
5000Vswitchesbetween500V,1000V,2500V,5000V.
·Highvoltageindication
ThereisaredLEDlightinthemachineindicatingthatthereishighvoltageoutputinthe
instrument,thereisadangerofelectricshock,theoperatorshouldpayspecialattentionto
safety.
·Lowbatteryvoltageindication"".
·Battery-driven,easiertouse.
·2000digitswithbargraphdisplay,easytoreadthedata.
Measuringrange:
1000V:
0.1MΩ20GΩ~
2500V:1MΩ~200GΩ
5000V:10M~200GΩ,automaticrangeconversion.
·Easytooperateandcarry.
·Strongloadcapacity,<1.8mAoutputshort-circuitcurrent.
·Completeprotectioncircuit,notafraidofvoltagecounterattack,outputshort
circuitprotection.
·ItcanmeasureAC/DCvoltage10V-600V,ACvoltagefrequency40Hz~70Hz.
·Dust-proofandmoisture-proofstructure,suitableforoutdooroperation.
·Dimensions:180X140X70(mm)
·Weight:about800g(includingbattery)
A
.TechnicalData
Workinghumidityconditions:0°C~35°C,relative 75%
orless,2000metersabovesea
level
Outputvoltage:(output±10% load>=1000M)Ω
Short
circuitcurrent: aboutlessthan1.8mA;
Lowbatteryindication:Displayshows
Insulationresistance:>=500MΩ(1000V);
Withstandvoltage:AC2KV50Hz1minute;
Working
temperatureandworkinghumidity: temperature0°C~40°C,relative humidity
85%orless;
Table2.2500Vtechnicalparameters
Outputvoltage:±10%
250V/500V/1000V/2000V/2500V
B
.Appearance
1.InsulationresistancemeasurementhighvoltageoutputjackL
2.PositiveterminalofAC/DCvoltagemeasurement
3.VoltagemeasurementinputnegativeshieldingendG(shieldingend)
4.InsulationresistancemeasurementsamplingjackE
Produktspecifikationer
| Varumärke: | PCE Instruments |
| Kategori: | Mätutrustning |
| Modell: | PCE-MO 1500 |
Behöver du hjälp?
Om du behöver hjälp med PCE Instruments PCE-MO 1500 ställ en fråga nedan och andra användare kommer att svara dig
Mätutrustning PCE Instruments Manualer
29 Mars 2025
28 Mars 2025
28 Mars 2025
28 Mars 2025
28 Mars 2025
28 Mars 2025
28 Mars 2025
28 Mars 2025
28 Mars 2025
28 Mars 2025
Mätutrustning Manualer
Nyaste Mätutrustning Manualer
3 April 2025
3 April 2025
3 April 2025
3 April 2025
3 April 2025
3 April 2025
3 April 2025
3 April 2025
3 April 2025
3 April 2025