Rossmax SD100 Bruksanvisning
                    Rossmax
                    
                    Pulsoximetrar
                    
                    SD100
                
                                
    Läs gratis den bruksanvisning för Rossmax SD100 (10 sidor) i kategorin Pulsoximetrar. Guiden har ansetts hjälpsam av 26 personer och har ett genomsnittsbetyg på 4.3 stjärnor baserat på 13.5 recensioner. Har du en fråga om Rossmax SD100 eller vill du ställa frågor till andra användare av produkten? Ställ en fråga
                        Sida 1/10
                    
                    
                    
Cleaning
Maintenance
Troubleshooting
Specification 
1.  Please clean the surface of the device before using. 
Wipe the device with medical alcohol (70% (w/w) 
Ethanol) first, and then let it dry in air or clean it by 
dry clean fabric. When cleaning the device with 
water, the water temperature should be lower than 
60°C
2.  Using the medical alcohol to disinfect the product 
after use, prevent from cross infection for next time 
use.
3.  The best storage environment of the device is - 25°C 
to 70°C ambient temperature and not higher than 
90% relative humidity. 
Note:  1.  Do not sterilize, autoclave or immerse this de-
vice in liquid. Do not pour or spray any liquids 
onto the device.
            2.  Do  not  use  caustic  or  abrasive  cleaning 
agents, or any cleaning agent containing am-
monium chloride or isopropyl alcohol.
Recommends user to return this device to the man-
ufacturer perform the following checks every 24 
months.
-  Inspect the equipment for mechanical and function-
al damage or deterioration.
-  Ensure all user interface keys and accessories func-
tion normally.
Note:  Manufacturer use Index2 SpO2 simulator to 
verify operation of the pulse oximeter device.
Symptoms Check points Corrections
SpO2 or 
pulse rate 
cannot 
displayed
Applied nger 
improperly.
Place the nger prop-
erly and try again.
SpO2 is too low to 
detect
Try again; go to con-
sult with your physi-
cian if you are sure the 
device works well.
SpO2 or 
pulse rate are 
not displayed 
stably
Applied nger 
improperly.
Place the nger prop-
erly and try again.
Finger is shaking 
or body is moving.
keep body steady
No display 
when button 
is pressed
Batteries run 
down
Replace with new 
batteries
Batteries not 
inserted correctly.
Re-insert batteries
The display 
disappears 
suddenly
The device will 
auto power o 
when it gets no 
signal.
Normal
Low battery
Replace with new 
batteries
 Note:  If the unit does not work, return it to your 
dealer. Under no circumstance should you 
disassemble and repair the unit by yourself.
SpO2 
Measuring 
range
35%~99%, (the resolution is 1%).
Accuracy 70%~99%: ±2%, Below 35~69%: 
unspecied.
Optical Sensor The wavelength of red LED is 660 nm 
and Infrared LED is 905/880 nm with 
maximum optical output power of 4 
mW/sr.
Pulse 
Measuring 
range
30 bpm~250 bpm (the resolution is 1 
bpm)
Accuracy ±3 bpm
Power source AAA × 2 (Alkaline)
Battery life Continually for 16 hours with two alkaline 
batteries
Operating 
Condition
Temperature: 5°C~35°C (41°F ~ 95°F), 
Relative Humidity: 15-95% (non 
condensing), Atmospheric pressure: 
700hPa ~ 1060hPa, Attitude: -1,280 to 
12,000 feet (-390m to 3,658m)
Storage / 
Transportation 
Condition
Temperature: -25°C~+70°C(-13°F ~ 
158°F), Relative humidity: 15-90%(non 
condensing), Atmospheric pressure: 
700hPa ~ 1060hPa, Attitude: -1,280 to 
12,000 feet (-390m to 3,658m), The time 
from 70°C or -25°C back to use : 3 hours
Dimensions 61.3(L) × 31W) × 36.7(H) mm
Weight About 33g (without the batteries)
Standards IEC60601-1-2, Class B, IEC60601-1, Type BF, 
ISO80601-2-61, IEC/EN60601-1-11
Type BF applied parts
IP Classication IP22: Protection against harmful ingress of 
water and particulate matter
Note: 
-  A description of the effect on displayed and transmit-
ted SpO2 and pulse rate:
-  Data averaging: 4 seconds for SpO2; 8 seconds for 
pulse rate.
- Data update delay: Less than 2 seconds.
be evaluated at the measurement site. Rather, they 
are exclusively used to display the current signal vari-
ation at the measurement site and do not enable re-
liable diagnostics for the pulse.
-  A warning that other cables and accessories may 
negatively aect EMC performance.
-  A warning regarding stacking and location close to 
other equipment.
-  A warning that use of other accessories results in 
non-compliance.
-  The safety way in high-temperature environment for 
all people use is measuring for 10 minutes, and turn 
it o for 20 minutes before measure again.
-  The oximeter is calibrated in the factory before sale, 
there is no need to calibrate it during its life cycle.
Model: SD100
 www.rossmax.com
 Fingertip Pulse Oximeter
Rossmax InnoTek Corp.  
12F., No. 189, Kang Chien Rd., Taipei, 
114, Taiwan.
CMC Medical Devices & Drugs S.L.
C/ Horacio Lengo Nº 18, CP 29006, 
Málaga, Spain
   WARNING: The symbol on this product means that it's an electronic product 
and following the European directive 2012/19/EU the electronic products have 
to be dispose on your local recycling centre for safe treatment.
EMC guidance and manufacturer’s declaration
Recommended separation distances between portable and mobile RF communications equipment and the ME equipment
The Finger-tip pulse oximeter is intended for use in an electromagnetic environment in which radiated RF disturbances are con-
trolled. The customer or the user of the Finger-tip pulse oximeter can help prevent electromagnetic interference by maintaining a 
minimum distance between portable and mobile RF communications equipment (transmitters) and the Finger-tip pulse oximeter 
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output 
power of transmitter / W
Separation distance according to frequency of transmitter / m
150 kHz to 80 MHz , d=[3.5/
V1]√P
80 MHz to 800 MHz , d=[3.5/
E1]√P
800 MHz to 2,5 GHz , d=[7/
E1]√P
0.01 0.1 0.1 0.2
0.1 0.4 0.4 0.7
1 1.2 1.2 2.3
10 3.7 3.7 7.4
100 11.7 11.7 23.3
Declaration – electromagnetic emissions
The Finger-tip pulse oximeter is intended for use in the electromagnetic environment specied below. The customer or the user of 
the Finger-tip pulse oximeter should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11 Group 1 Portable and mobile RF communications equipment should be used no 
closer to any part of the EQUIPMENT or SYSTEM including cables, than the 
recommended separation distance calculated from the equation applicable 
to the frequency of the transmitter. Interference may occur in the vicinity of 
equipment marked with the following symbol.  
RF emissions CISPR 11 Class B The Finger-tip pulse oximeter is suitable for use in all establishments, 
including domestic establishments and those directly connected to the 
public low-voltage power supply network that supplies buildings used 
for domestic purposes.
Harmonic emissions IEC 61000-3-2 N/A
Voltage uctuations/Flicker emissions 
IEC 61000-3-3
N/A
Declaration – electromagnetic emissions and immunity – for EQUIPMENT and SYSTEMS that are use in the professional health-
care facility environment or in the home healthcare environment
The Finger-tip pulse oximeter declaration – electromagnetic immunity
The Finger-tip pulse oximeter system is intended for use in the electromagnetic environment specied below. The customer or 
the user of the Finger-tip pulse oximeter system should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Conducted RF IEC 
61000-4-6
3 Vrms ; 6 Vrms ; 150 
kHz to 80 MHz
N/A Portable and mobile RF communications equipment 
should be used no closer to any part of the EQUIPMENT 
or SYSTEM including cables, than the recommended 
separation distance calculated from the equation ap-
plicable to the frequency of the transmitter. Interference 
may occur in the vicinity of equipment marked with the 
following symbol. 
Radiated RF IEC 
61000-4-3
3 V/m ; 10V/m ; 80 
MHz – 2.7 GHz l 80%
3 V/m ; 10V/m ; 80 
MHz – 2.7 GHz ; 80%
Proximity elds 
from RF wireless 
Communications 
equipment IEC 
61000-4-3
27 V/m 385 MHz 27 V/m 385 MHz
28 V/m 450 MHz 28 V/m 450 MHz
9 V/m 710 MHz 9 V/m 710 MHz
745 MHz 745 MHz
780 MHz 780 MHz
28 V/m 810 MHz 28 V/m 810 MHz
870 MHz 870 MHz
930 MHz 930 MHz
28 V/m 1720 MHz 28 V/m 1720 MHz
1845 MHz 1845 MHz
1970 MHz 1970 MHz
28 V/m 2450 MHz 28 V/m 2450 MHz
9 V/m 5240 MHz 9 V/m 5240 MHz
5500 MHz 5500 MHz
5785 MHz 5785 MHz
Declaration – electromagnetic immunity
The Finger-tip pulse oximeter system is intended for use in the electromagnetic environment specied below. The customer or 
the user of the Finger-tip pulse oximeter system should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge 
(ESD) IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, 
±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 
kV, ±15 kV air
Floors should be wood, concrete or ceramic tile. 
If oors are covered with synthetic material, the 
relative humidity should be at least 30 %.
Electrical fast transient/
burst IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
N/A Mains power quality should be that of a typical 
commercial or hospital environment.
Surge IEC 61000-4-5 ±0.5 kV
±1 kV dierential mode
±2 kV common mode
N/A Mains power quality should be that of a typical 
commercial or hospital environment.
Voltage dips, short 
interruptions and voltage 
variations on power 
supply input lines IEC 
61000-4-11
0 % U
T
; 0, 5 cycle At 0°, 45°, 
90°, 135°, 180°, 225°, 270° 
and 315°
0 % U
T
; 1 cycle And 70 
% U
T
; 25/30 cycle Single 
phase: at 0°
N/A Mains power quality should be that of a typical 
commercial or hospital environment. If the user 
of the EQUIPMENT or SYSTEM requires continued 
operation during power mains interruptions, it is 
recommended that the EQUIPMENT or SYSTEM 
be powered from an uninterruptible power sup-
ply or a battery.
Power frequency (50/60 
Hz) magnetic eld IEC 
61000-4-8
30 A/m 30 A/m Power frequency magnetic elds should be at 
levels characteristic of a typical location in a typi-
cal commercial or hospital environment.
Warranty Card
This instrument is covered by a 2 years guarantee from the date of purchase, 
batteries and accessories are not included. The guarantee is valid only on 
presentation of the guarantee card completed by the dealer confirming date 
of purchase or the receipt. Opening or altering the instrument invalidates 
the guarantee. The guarantee does not cover damage, accidents or non-
compliance with the instruction manual. Please contact your local seller/dealer 
or www.rossmax.com.
Customer Name:  _______________________________
Address:  ____________________________________
Telephone:  __________________________________
E-mail address:  ________________________________
Product Information:  ___________________________
Date of purchase: ______________________________
Store where purchased: 
___________________________________________
IN0SD100000000011
RI_IB_SD100_EN_
TP_ver2214
Produktspecifikationer
| Varumärke: | Rossmax | 
| Kategori: | Pulsoximetrar | 
| Modell: | SD100 | 