CALIBRATION PROCEDURE FOR SCALE PZC-5/10
1. PUT THE SCALE IN A PLACE FREE OF ANY MOVEMENT OR VIBRATION.
2. MAKE SURE THAT THE SCALE IS LEVELED AND WELL SUPPORTED ON ITS
FOUR BASES, CONNECT THE SCALE TO A 120 VAC OUTLET.
3. DURING THE COUNTDOWN, PRESS THE KEYS, KEYS IN ,, ,
THIS SEQUENCE.
4. AFTER THE INITIAL COUNTDOWN, THE DISPLAY WILL READ “CODE”. ENTER
THE CALIBRATION PASSWORD BY PRESSING THE FOLLOWING KEYS:
PASSWORD FOR LB: 100101
= 0 = 1
AFTER ENTERING THE PASSWORD PRESS THE KEY.
5. IF THE PASSWORD IS CORRECT, THE INDICATOR WILL READ “REF ZERO”.
6. WITH NOTHING ON THE SCALE PRESS THE KEY. THIS WILL BE THE
SCALES REFERENCE TO ZERO.
6. NOW THE SCALE WILL READ “LOAD”. PLACE 10 LB ON THE SCALE AND
PRESS THE KEY.
7. WAIT UNTIL YOU HEAR A BEEP AND THE CALIBRATION COEFFICIENT
APPEARS ON THE INDICATOR. AFTERWARDS, REMOVE THE WEIGHT FROM
THE PLATE AND TURN THE SCALE OFF. THE SCALE SHOULD NOW BE
CALIBRATED
8. VERIFY THE ACCURACY IN REPEAT THIS PROCEDURE IF NEEDED.AND
NOTE: THE CALIBRATION PASSWORD CHANGES ACCORDING TO THE
SERIAL NUMBER.