Läs gratis den bruksanvisning för Klein Tools MM5000N (8 sidor) i kategorin Multimeter. Guiden har ansetts hjälpsam av 2 personer och har ett genomsnittsbetyg på 4.3 stjärnor baserat på 1.5 recensioner. Har du en fråga om Klein Tools MM5000N eller vill du ställa frågor till andra användare av produkten? Ställ en fråga

Sida 1/8
ENGLISH
Instruction
Manual
ENGLISH
MM5000
• TRUERMS
• AUTO/MANUAL
RANGE
• AUTOHOLD
• MAX/MIN
• BARGRAPH
• BACKLITDISPLAY
• 3-3/4DIGIT
4000COUNTLCD
• LEADSTORAGE
4007036
ELECTRICALSPECIfICATIONS
DCVoltageMeasurement
Range Resolution Accuracy
400.0mV 0.1mV
± (0.3% + 5 digits)
4V ~ 400V 1mV ~ 0.1V
1000V 1V ± (0.8% + 10 digits)
OverloadProtection: 1000V InputImpedance(Nominal): >10MΩ, <100pF
InputImpedance(LOWZ): >3kΩ, <200pF, up to 600V RMS
ACVoltageMeasurement
Range Resolution
Accuracy
40Hz ~ 400Hz
Accuracy
400Hz ~ 2kHz
400mV 0.1mV
± (0.75% + 5 digits)
± (2.5% +10 digits)
4V ~ 400V 1mV ~ 0.1V ± (2.0% +8 digits)
1000V 1V ± (0.75% + 8 digits) ± (2.0% +8 digits)*
OverloadProtection: 1000V RMS InputImpedance(Nominal): >10MΩ, <100pF
InputImpedance(LOWZ): >3kΩ, <200pF, up to 600V RMS
frequencyRange: 40 to 2kHz Response: True RMS
*Accuracy for 400Hz ~ 1kHz range only.
DCCurrentMeasurement
Range Resolution Accuracy
400.0µA 0.1µA
± (1.0% + 5 digits)
2000µA 1µA
OverloadProtection: 600V RMS
ACCurrentMeasurement
Range Resolution Accuracy
400.0µA 0.1µA
± (1.5% + 5 digits)
2000µA 1µA
OverloadProtection: 600V RMS
frequency: 40Hz to 2kHz Response: True RMS
ResistanceMeasurement
Range Resolution Accuracy
400Ω
˜
4MΩ 0.1Ω
˜
1kΩ ± (1.0% + 5 digits)
40MΩ 0.01MΩ ± (1.5% + 10 digits)
OverloadProtection: 600V RMS
DiodeTest
OverloadProtection Range TestCurrent OpenCircuitVoltage
600V RMS 4.0V DC Appx. 0.25mA
<
1.8V DC (typ.)
ContinuityTest
OverloadProtection OpenCircuitVoltage Threshold(Appx.)
600V RMS Appx. 0.44V Appx. < 40Ω
Specicationssubjecttochange.
WARRANTY
www.kleintools.com/warranty
CLEANING
Turn instrument off and disconnect test leads. Clean the instrument by using a
damp cloth. Do not use abrasive cleaners or solvents.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period
of time. Do not expose to high temperatures or humidity. After a period of
storage in extreme conditions exceeding the limits mentioned in the General
Specifications section, allow the instrument to return to normal operating
conditions before using it.
DISPOSAL/RECYCLE
Do not place equipment and its accessories in the trash.
Items must be properly disposed of in accordance with local
regulations
.
Prior to disposal of this product, please contact Klein Tools
for proper disposal options.
CUSTOMERSERVICE
KLEINTOOLS,INC.
450 Bond Street
Lincolnshire, IL 60069
1-877-775-5346
www.kleintools.com
customerservice@kleintools.com
Rev 04/14 B 139749
7. LowImpedance(LOWZ):<600V
DONOTuse(LOWZ)modeatvoltagesgreaterthan600V.
• Press SELECT to choose AC or DC Voltage.
• Use Low Impedance (LOW Z) mode to reduce ghost voltages.
• See AC/DC Voltage instructions.
6. Continuity
Features:
HOLD
MAX/MIN
• Display shows resistance.
• Buzzer sounds if less than 40Ω.
5. DiodeTest
Features:
HOLD
MAX/MIN
Display shows:
• Forward voltage drop if forward biased.
• “O.L” if reverse biased.
4. AC/DCCurrent(small):<2000µA
Features:
HOLD
AUTOHOLD
RANGE
MAX/MIN
DONOTattempttomeasuremorethan400mA.
• Attach RED lead to “µA” input.
• Break circuit to be measured.
• Attach RED lead to "µA" input, BLACK lead to COM.
• Connect RED lead to one side of break and BLACK lead to the
other (meter in series).
• Select AC or DC current source.
Dwg Name: MM5000-139749ART Dwg No: 139749
ECO No: 016671 Pkg Dwg Ref: 1650 Rev: B
Color Reference: CEN-PACK-003-A
MM5000-139749-Instructions.indd 1-4 5/5/14 10:14 AM
ENGLISH

MM5000


The Klein Tools MM5000 is an auto-ranging multimeter. It measures AC/DC
voltage, low impedance voltage, AC/DC current and resistance. It can also
test diodes and continuity.
• 6,562 ft. / 2,000 m
• <75%
• 32°F to 113°F (0°C to 45°C)
• -4°F to 140°F (-20°C to 60°C ) <80% R.H.
• 18°C to 28°C ( 64°F to 82°F)
•  0.1*(specified accuracy)/ °C
• 3 samples per second
•  7” x 3.5” x 1.875” (178 x 89 x 48 mm)
• 14 oz. (397 g)
• 160 Hours
•  3 m
•  40
• Accurate for one year
• ± (% of reading + # of least significant digits)
• CAT III 1000V, CAT IV 600V
• Conforms to UL Stds 61010-1 & 61010-2-030.
Certified to CAN/CSA C22.2 No.61010-1, 61010-2-030,
IEC 61010-2-030:2010 MOD, ANSI/ISA-61010-2-030 (82.02.03).
• 2
• USA (of US and imported parts)

To ensure safe operation and service of the tester, follow these instructions.
Failure to observe these warnings can result in severe injury or death.
• Before each use, verify meter operation by measuring a known
voltage or current.
• Never use the meter on a circuit with voltages that exceed the
category based rating of this meter.
• Do not use the meter during electrical storms or in wet weather.
• Do not use the meter or test leads if they appear to be damaged.
• Ensure meter leads are fully seated, and keep fingers away from the
metal probe contacts when making measurements.
• Do not open the meter to replace batteries or fuses while the probes
are connected.
• Use caution when working with voltages above 60V DC or 25V AC
RMS. Such voltages pose a shock hazard.
• To avoid false readings that can lead to electrical shock, replace
batteries if a low battery indicator appears.
• Unless measuring voltage or current, shut off and lock out power
before measuring resistance or capacitance.
• Always adhere to local and national safety codes. Use individual
protective equipment to prevent shock and arc blast injury where
hazardous live conductors are exposed.

~
 
 
 
 
 


 
• Switch between AC and DC.
• Wake meter during Auto Power Off.
 
• Hold captures the current displayed value.
• Auto Hold captures the first stable displayed value, until a new
stable value is measured. The meter will then capture the new value
and emit a beep (V, Ohm and Amp functions).
• Press to hold current input on the display.
• Press again to enter Auto-Hold to capture stable values.
• Press again to return to normal operation.
 
• Press repeatedly to cycle through manual ranges.
• Press for 2 seconds to return to auto ranging mode.
• AT is displayed on LCD only during auto ranging mode.
 
• Press to enter Max/Min mode; the largest and smallest values will
be saved while in this mode.
• Press repeatedly to alternate between the maximum and minimum
readings.
• Press for 2 seconds to return to live reading and clear the stored
maximum and minimum values.
 
• Device will power off after 30 minutes non-use.
• Press the select button to wake.
• Disabled during Max/Min function.
• Holding Select button while turning on disables Auto Power-Off.
 
• When indicator is displayed on the LCD, batteries must be
replaced.
• Remove rubber boot, back screw and replace 2 x AAA batteries.
 
• Press to activate display backlight.
• Using lights drains the battery significantly.
 
• The bar graph shows an approximate analog representation
of a measurement.
• The bar graph responds much faster than the digital display.
• The scale of the bar graph is zero to the maximum reading of
the selected range.
 (optional, sold separately) 
• Slide magnetic adapter into protective rubber boot.
• Attach instruments to metal for hands-free use.

 
Features:


RANGE



• Use Low Impedance (LOW Z) mode to reduce ghost voltages.
• Rotate selector to "AC" Voltage position.
• Attach RED lead to “V” input.
 
Features:


RANGE



• Use Low Impedance (LOW Z) mode to reduce ghost voltages.
• Rotate selector to "DC" Voltage position.
• Attach RED lead to “V” input.
A
B
C
E
G
H
I
K
L
J
3
1
5
7
 measure resistance on a live circuit.
• Ω = < 40MΩ.

~


-

AT

O
.L.

Apo




MIN

M
AX

V

A

Ω
 

n


µ


m

-3
k

3
M

A- HOLD

LOW
Z

 
Features:


RANGE

Dwg Name: MM5000-139749ART Dwg No: 139749
ECO No: 016671 Pkg Dwg Ref: 1650 Rev: B
Color Reference: CEN-PACK-003-A
MM5000-139749-Instructions.indd 5-8 5/5/14 10:14 AM