Läs gratis den bruksanvisning för Klein Tools MM5000N (8 sidor) i kategorin Multimeter. Guiden har ansetts hjälpsam av 17 personer och har ett genomsnittsbetyg på 4.2 stjärnor baserat på 9 recensioner. Har du en fråga om Klein Tools MM5000N eller vill du ställa frågor till andra användare av produkten? Ställ en fråga

Sida 1/8
ENGLISH
Instruction
Manual
ENGLISH
MM5000
• TRUERMS
• AUTO/MANUAL
RANGE
• AUTOHOLD
• MAX/MIN
• BARGRAPH
• BACKLITDISPLAY
• 3-3/4DIGIT
4000COUNTLCD
• LEADSTORAGE
4007036
ELECTRICALSPECIfICATIONS
DCVoltageMeasurement
Range Resolution Accuracy
400.0mV 0.1mV
± (0.3% + 5 digits)
4V ~ 400V 1mV ~ 0.1V
1000V 1V ± (0.8% + 10 digits)
OverloadProtection: 1000V InputImpedance(Nominal): >10MΩ, <100pF
InputImpedance(LOWZ): >3kΩ, <200pF, up to 600V RMS
ACVoltageMeasurement
Range Resolution
Accuracy
40Hz ~ 400Hz
Accuracy
400Hz ~ 2kHz
400mV 0.1mV
± (0.75% + 5 digits)
± (2.5% +10 digits)
4V ~ 400V 1mV ~ 0.1V ± (2.0% +8 digits)
1000V 1V ± (0.75% + 8 digits) ± (2.0% +8 digits)*
OverloadProtection: 1000V RMS InputImpedance(Nominal): >10MΩ, <100pF
InputImpedance(LOWZ): >3kΩ, <200pF, up to 600V RMS
frequencyRange: 40 to 2kHz Response: True RMS
*Accuracy for 400Hz ~ 1kHz range only.
DCCurrentMeasurement
Range Resolution Accuracy
400.0µA 0.1µA
± (1.0% + 5 digits)
2000µA 1µA
OverloadProtection: 600V RMS
ACCurrentMeasurement
Range Resolution Accuracy
400.0µA 0.1µA
± (1.5% + 5 digits)
2000µA 1µA
OverloadProtection: 600V RMS
frequency: 40Hz to 2kHz Response: True RMS
ResistanceMeasurement
Range Resolution Accuracy
400Ω
˜
4MΩ 0.1Ω
˜
1kΩ ± (1.0% + 5 digits)
40MΩ 0.01MΩ ± (1.5% + 10 digits)
OverloadProtection: 600V RMS
DiodeTest
OverloadProtection Range TestCurrent OpenCircuitVoltage
600V RMS 4.0V DC Appx. 0.25mA
<
1.8V DC (typ.)
ContinuityTest
OverloadProtection OpenCircuitVoltage Threshold(Appx.)
600V RMS Appx. 0.44V Appx. < 40Ω
Specicationssubjecttochange.
WARRANTY
www.kleintools.com/warranty
CLEANING
Turn instrument off and disconnect test leads. Clean the instrument by using a
damp cloth. Do not use abrasive cleaners or solvents.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period
of time. Do not expose to high temperatures or humidity. After a period of
storage in extreme conditions exceeding the limits mentioned in the General
Specifications section, allow the instrument to return to normal operating
conditions before using it.
DISPOSAL/RECYCLE
Do not place equipment and its accessories in the trash.
Items must be properly disposed of in accordance with local
regulations
.
Prior to disposal of this product, please contact Klein Tools
for proper disposal options.
CUSTOMERSERVICE
KLEINTOOLS,INC.
450 Bond Street
Lincolnshire, IL 60069
1-877-775-5346
www.kleintools.com
customerservice@kleintools.com
Rev 04/14 B 139749
7. LowImpedance(LOWZ):<600V
DONOTuse(LOWZ)modeatvoltagesgreaterthan600V.
• Press SELECT to choose AC or DC Voltage.
• Use Low Impedance (LOW Z) mode to reduce ghost voltages.
• See AC/DC Voltage instructions.
6. Continuity
Features:
HOLD
MAX/MIN
• Display shows resistance.
• Buzzer sounds if less than 40Ω.
5. DiodeTest
Features:
HOLD
MAX/MIN
Display shows:
• Forward voltage drop if forward biased.
• “O.L” if reverse biased.
4. AC/DCCurrent(small):<2000µA
Features:
HOLD
AUTOHOLD
RANGE
MAX/MIN
DONOTattempttomeasuremorethan400mA.
• Attach RED lead to “µA” input.
• Break circuit to be measured.
• Attach RED lead to "µA" input, BLACK lead to COM.
• Connect RED lead to one side of break and BLACK lead to the
other (meter in series).
• Select AC or DC current source.
Dwg Name: MM5000-139749ART Dwg No: 139749
ECO No: 016671 Pkg Dwg Ref: 1650 Rev: B
Color Reference: CEN-PACK-003-A
MM5000-139749-Instructions.indd 1-4 5/5/14 10:14 AM
ENGLISH

MM5000


The Klein Tools MM5000 is an auto-ranging multimeter. It measures AC/DC
voltage, low impedance voltage, AC/DC current and resistance. It can also
test diodes and continuity.
• 6,562 ft. / 2,000 m
• <75%
• 32°F to 113°F (0°C to 45°C)
• -4°F to 140°F (-20°C to 60°C ) <80% R.H.
• 18°C to 28°C ( 64°F to 82°F)
•  0.1*(specified accuracy)/ °C
• 3 samples per second
•  7” x 3.5” x 1.875” (178 x 89 x 48 mm)
• 14 oz. (397 g)
• 160 Hours
•  3 m
•  40
• Accurate for one year
• ± (% of reading + # of least significant digits)
• CAT III 1000V, CAT IV 600V
• Conforms to UL Stds 61010-1 & 61010-2-030.
Certified to CAN/CSA C22.2 No.61010-1, 61010-2-030,
IEC 61010-2-030:2010 MOD, ANSI/ISA-61010-2-030 (82.02.03).
• 2
• USA (of US and imported parts)

To ensure safe operation and service of the tester, follow these instructions.
Failure to observe these warnings can result in severe injury or death.
• Before each use, verify meter operation by measuring a known
voltage or current.
• Never use the meter on a circuit with voltages that exceed the
category based rating of this meter.
• Do not use the meter during electrical storms or in wet weather.
• Do not use the meter or test leads if they appear to be damaged.
• Ensure meter leads are fully seated, and keep fingers away from the
metal probe contacts when making measurements.
• Do not open the meter to replace batteries or fuses while the probes
are connected.
• Use caution when working with voltages above 60V DC or 25V AC
RMS. Such voltages pose a shock hazard.
• To avoid false readings that can lead to electrical shock, replace
batteries if a low battery indicator appears.
• Unless measuring voltage or current, shut off and lock out power
before measuring resistance or capacitance.
• Always adhere to local and national safety codes. Use individual
protective equipment to prevent shock and arc blast injury where
hazardous live conductors are exposed.

~
 
 
 
 
 


 
• Switch between AC and DC.
• Wake meter during Auto Power Off.
 
• Hold captures the current displayed value.
• Auto Hold captures the first stable displayed value, until a new
stable value is measured. The meter will then capture the new value
and emit a beep (V, Ohm and Amp functions).
• Press to hold current input on the display.
• Press again to enter Auto-Hold to capture stable values.
• Press again to return to normal operation.
 
• Press repeatedly to cycle through manual ranges.
• Press for 2 seconds to return to auto ranging mode.
• AT is displayed on LCD only during auto ranging mode.
 
• Press to enter Max/Min mode; the largest and smallest values will
be saved while in this mode.
• Press repeatedly to alternate between the maximum and minimum
readings.
• Press for 2 seconds to return to live reading and clear the stored
maximum and minimum values.
 
• Device will power off after 30 minutes non-use.
• Press the select button to wake.
• Disabled during Max/Min function.
• Holding Select button while turning on disables Auto Power-Off.
 
• When indicator is displayed on the LCD, batteries must be
replaced.
• Remove rubber boot, back screw and replace 2 x AAA batteries.
 
• Press to activate display backlight.
• Using lights drains the battery significantly.
 
• The bar graph shows an approximate analog representation
of a measurement.
• The bar graph responds much faster than the digital display.
• The scale of the bar graph is zero to the maximum reading of
the selected range.
 (optional, sold separately) 
• Slide magnetic adapter into protective rubber boot.
• Attach instruments to metal for hands-free use.

 
Features:


RANGE



• Use Low Impedance (LOW Z) mode to reduce ghost voltages.
• Rotate selector to "AC" Voltage position.
• Attach RED lead to “V” input.
 
Features:


RANGE



• Use Low Impedance (LOW Z) mode to reduce ghost voltages.
• Rotate selector to "DC" Voltage position.
• Attach RED lead to “V” input.
A
B
C
E
G
H
I
K
L
J
3
1
5
7
 measure resistance on a live circuit.
• Ω = < 40MΩ.

~


-

AT

O
.L.

Apo




MIN

M
AX

V

A

Ω
 

n


µ


m

-3
k

3
M

A- HOLD

LOW
Z

 
Features:


RANGE

Dwg Name: MM5000-139749ART Dwg No: 139749
ECO No: 016671 Pkg Dwg Ref: 1650 Rev: B
Color Reference: CEN-PACK-003-A
MM5000-139749-Instructions.indd 5-8 5/5/14 10:14 AM