Rossmax NH60 Bruksanvisning


Läs gratis den bruksanvisning för Rossmax NH60 (2 sidor) i kategorin Medicinsk. Guiden har ansetts hjälpsam av 29 personer och har ett genomsnittsbetyg på 4.9 stjärnor baserat på 15 recensioner. Har du en fråga om Rossmax NH60 eller vill du ställa frågor till andra användare av produkten? Ställ en fråga

Sida 1/2
1
1.0 Object
1.1
To test the performance of Rossmax NH60 nebulizer
2.0 Equipment List
2.1 Rossmax NH60 Nebulizer*2
2.2 Rossmax Nebulizer kit*2
2.3 Malvern Spraytec particle size analyer
2.4 Marple 298 Cascade Impactor
2.5 Chroma 61602 Programmable AC Source
2.6 Shimadzu AUW120D microbalance
2.7 A.P. Buck, Inc. Libra Plus LP-5 sampling pump
2.8 SSI P51-6BarS-A-MD-20mA pressure meter
2.9 Golden Mountain Enterprise Co. Ltd. F33L0096 flow meter
2.10 Humidity/Temperature Meter
2.11 Taiwan Biotech Co., Ltd 0.9% Saline solution
2.12 2.5% NaF solution
2.13 Atrovent Ipratropium Bromide
2.14
Atrovent Flixotide
2.15 AstraZeneca Terbutaline Sulphate
2.16 Ventoline (2.5mg) Salbutamol/Sulphate
2.17 Casio Timer
3.0 Testing Items
3.1 Aerosol Particle Size Distribution Testing(By Malvern Spraytec)
3.2 Aerosol Particle Size Distribution Testing(By Marple 298 Cascade Impactor)
3.3 Nebulization Rate Testing(Including drugs testing)
3.4 Residual Volume Testing
3.5 Reliability Test
4.0 Testing Procedure
4.1 Aerosol Particle Size Distribution Testing(By Malvern Spraytec)
4.1.1 Each sample should be tested with 2.5ml 0.9% saline solution for 3 minutes.
4.1.2 Add 2.5ml 0.9% saline solution into the nebulizer kit,
4.1.3 Connect the nebulizer kit with NH60 and put at the testing site, the nebulizer kit’s
outlet must be kept at 3.0 cm from the laser beam.
4.1.4 Start recording Spraytec for more than 15 secs, then start NH60 for testing.
4.1.5 After 3.0 minutes have been reached, stop the NH60 and then stop Spraytec.
4.1.6 Checks Spraytec records