YSI Incorporated
5400 & 5500D-01, 5500D-02, 5500D-04
MultiDO
Monitor, Control, & Alarm Instruments
User Manual
YSI 5X00
2
Item # 605877
Rev F
Drawing # A605877
May 2015
Copyright
©2015 YSI Incorporated.
Manufacturer:
YSI, Incorporated
A Xylem Brand
1700/1725 Brannum Lane
Yellow Springs, OH 45387 USA
937-767-7241 www.ysi.com
All trademarks herein are
registered to YSI Incorporated
unless otherwise noted.
Teon is a trademark of DuPont.
Microso and Windows are registered trademarks of the Microso
Corporation in the United States and other countries.
Pentium is a registered trademark of Intel Corporation.
Document information
YSI 5X00
4
Table of Contents
Directions to Wire AC Power 45 ..............................................
Wire DC Power and Backup Power 46 .........................................
Directions to Wire DC Power/DC Backup Power 47 ...............
Step 5 Wire Sensor(s) 49 ....................................................................
Wire DO Sensors 49 .....................................................................
Directions to Wire DO Sensor(s) 49 ........................................
Wire Temperature Sensor(s) ....................................................52
Directions to Wire Temperature Sensor(s) 53 .........................
Step 6 Wire Relay outputs 54 .............................................................
Directions to Wire Relay Outputs 54 .............................................
Wiring a Contactor 55 ...................................................................
Step 7 Wire Aux Inputs (Non Temperature) 56 ...................................
Directions to Wire Aux Inputs (Non Temperature Inputs) 57 .........
Step 8 Wire Network 60 ......................................................................
Directions to Wire Network 61 .......................................................
Step 9 Wire and Install Communication Equipment 66 .......................
Email Alarming .........................................................................66
RS232 Communication 66 ............................................................
Ethernet (TCP/IP) Module Communication 68 ..............................
Step 10 Install and Wire IOEM/REM instruments 68 ..........................
Lightning and Surge Protection 68 ......................................................
Sealants, Desiccants, and Securing 5X00 69 .....................................
4 Conguring 5X00 Systems ................................... 71
5X00 Conguration and Monitoring .....................................................72
Front Panel Keys 73 ...........................................................................
Softkeys 73 ...................................................................................
Run Screen 75 ....................................................................................
Run Screen Icons 76 .....................................................................
Menus 78 ............................................................................................
Submenu Screen Types ...........................................................80
Main Menu 81 .....................................................................................
Calibration Menu 81 ............................................................................
Getting Ready to Calibrate 82 .......................................................
Calibration Log 83 .........................................................................
DO 83 ............................................................................................
Directions to Calibrate DO Sensor %Sat 83 ............................
Directions to Calibrate DO Sensor mg/L ............................84
Local vs Remote Instruments 86 ........................................................
System Menu 86 .................................................................................
Adjust Contrast 87 .........................................................................
Buzzer 88 ......................................................................................
BackLighting Timer ...................................................................89
YSI 5X00
6
Validation Error and Autox - Set Point Control Only .........157
PID/PWM and PID 4/20mA Control mode 159 ..........................
Aux and E-Aux Relay Digital 164 ....................................................
Aux and E-Aux Analog (0-1VDC, 0-5VDC and 4/20mA) 166 .....
Temperature 168 .............................................................................
DO 1-4 173 ......................................................................................
Salinity & Temp Compensation for DO Measurements 173 ......
Two Range Controls for DO 173 ...............................................
Congure DO Systems .......................................................174
E-Aux system Ofine ..........................................................177
Congure sensor system as a 4/20mA output ....................178
AquaManager and the Sensor Setup Menu 179 .............................
Timers 180 ............................................................................................
Timers 1-4 180 ................................................................................
Directions to Congure Timers 1 - 4 ...................................180
Feed Smarttm Timers 1-4...........................................................182
Feed Mode 184 .........................................................................
Daily Weight ......................................................................185
Daily Feedings 185 ...................................................................
FCR 186 ...................................................................................
Sensors 187 ..............................................................................
Calibrate 189 .............................................................................
AquaManager and the Timer Menu 190 ..........................................
Control Menu 192 .................................................................................
Msg’s Menu 193 ....................................................................................
Calibration Logs 194 ......................................................................
Event Logs 194 ...............................................................................
Sensor Logs 195 .............................................................................
Svr Hold softkey Button 196 .................................................................
5 AquaManager 197 .........................................................
Installing AquaManager .......................................................................198
PC requirements and features 198 .......................................................
Software Installation 199 .......................................................................
AquaManager Windows 200 .................................................................
Explorer Window 200 ......................................................................
Node (Graph) Window 202 .............................................................
Systems Management 203 ........................................................
Database Management 204 ..................................................................
Autopoller 204 .......................................................................................
Mapping 205 ...................................................................................
AquaManager Alarm Emails .....................................................206
Table of Contents
YSI 5X00
8
Appendix 9 - Contact Us 258 ......................................................................
Ordering and Technical Support 258 ....................................................
Service Information 258 ........................................................................
Appendix 10 - Recycling 259 ......................................................................
Recycling 259 .......................................................................................
Battery Disposal 259 .............................................................................
Index .............................................................................. 260
Table of Contents
YSI 5X00
14
Environmental Specications
Environmental Category Operating Range
Temperature -15 to 70˚C (5 to 158˚F)
Enclosure Rating NEMA 4X (direct exposure to outside conditions) IP65
Compression Fitting Rating IP 68/NEMA 6 (liquid tight)
Electrical Safety IEC compliant*
Other RoHS compliant
See Appendix 4 Declaration of Conformity - 248.
Input System Specications
Aux Systems ere are 6 Aux input systems that allow digital or analog (0-
1.0VDC, 0-5.0VDC, or 4-20mA) conguration. In addition, four
of the Aux systems can be congured as temperature systems see
Temperature Systems Specs - 15.
Aux System Type Digital Analog
Range dry contact open/close 0 -1.0 VDC, 0-5.0 VDC, or 4-20mA
Accuracy n/a 0-1.0VDC: ± 3% ± 10 mV
0-5.0VDC: ± 3% ±15 mV
4-20 mA: ± 3% ±100μA
Resolution n/a user selectable (0), (0 .0), or (0.00)
DO Systems ere are up to 4 independent DO systems. Each DO system has
2 control ranges, 2 control ranges, and high and low alarming.
Sensor Type
5400
Galvanic (mV)
Membrane Type Teon (TE), 2 mil (gray)
Range 0 to 500% air saturation
0 to 60 mg/L
Accuracy ±2% or ± 0.2 mg/L, whichever is greater for 0-200%
±6% of reading for 200-500%
Resolution 0.1% (0.0-99.9%); 1% (> 99.9)
0.01 mg/L (0-10 mg/L); 0.1 mg/L (10-60 mg/L)
Introduction
YSI 5X00
18
is page le intentionally blank.
YSI 5X00
20
THIS IS IMPORTANT
Take time to be safe when working on electrical circuits and equipment. It could
save your life. Avoid unsafe acts and conditions. Do not operate or install the 5X00
before reading this chapter. Installing the 5X00 involves wiring high voltage equip-
ment. Follow all provided safety information. Installation must be performed by
someone who understand the dangers of, and is qualied to, wire electrical devices.
If information or instructions are not clear, DO NOT PROCEED until clarication
can be obtained.
ALWAYS:
Follow basic safety rules when working with or near high voltage circuits. e
5X00 is a high voltage instrument using 110v. Peripheral devices may also be
high voltage devices.
Consider the result of each act. Keep away from live circuits. Do not change
parts or make adjustments inside the equipment with high voltages on.
Do not service alone.
Do not tamper with interlocks. Safety devices such as interlocks, overload re-
lays, and fuses should never be altered or disconnected except for replacement.
Safety protection devices should never be changed or modied without a clear
understanding of what eect the modication will have on the system.
Do not ground yourself.
Do not energize equipment if there is any evidence of water leakage.
Warnings and Cautions
Safety information is presented in this manual in one of two ways:
- WARNING - RISK OF ELECTRIC SHOCK is symbol identies
installation and operation procedures that can result in personal injury
or death if safety information is not followed.
- AVERTISSEMENT RISQUE DE CHOC ÉLECTRIQUE. Ce symbole
identie les procédures d’installation et dopération pouvant entraîner des
blessures ou la mort si les consignes de sécurité ne sont pas suivies.
Safety
YSI 5X00
26
Hardware Installation
Correctly install 5X00 system components to ensure accurate data collection and
reliable operation. First, determine what monitor, control, and alarm functions
the 5X00 system will perform. Because system components are wired to the 5X00
instrument, it is important to determine where and how components will be installed
and congured. Preparing a system schematic (gure 3.1) is helpful.
Figure 3.1
Hardware installation steps (not all steps required for all applications):
Step 1 - install 5X00 instrument
Step 2 - locate and install sensor(s)
Step 3 - ground 5X00
Step 4 - wire power
Step 5 - wire sensors(s)
Step 6 - wire relays (output devices)
Step 7 - wire aux input devices (excluding temperature probes)
Step 8 - wire a network of 2 - 32 instruments
Step 9 - install and wire communication equipment
Step 10 - install and wire IOEM and REM instruments (optional). IOEM/REM
Installation information is included in a separate manual. See IOEM/REM
Installation and Safety Guide.
Installation and Wiring
YSI 5X00
28
• away from electromagnetic, radio, AC motor, transformer, or antennas
• so the front panel can be fully opened and serviced
• so system component wires are run as specied in this manual and according
to local applicable electrical code
Mount the 5X00
Select surface, rail, or panel mount option. Install the weather shield for outdoor
installations.
Surface Mount
Mount 5X00 to any at surface using mounting brackets. Mounting brackets and
screws are included with the 5X00.
1. Fasten the mounting brackets to the back of the 5X00 with the mounting
screws - gure 3.2.
2. Tighten the screws, securing the brackets to the back of the 5X00 enclosure.
3. Fasten the 5X00 mounting brackets to the mounting surface with the mount-
ing screws provided - gure 3.2. Make sure screws adequately secure 5X00
to surface.
4. Tighten the screws, securing the 5X00 to the surface.
Figure 3.2
Installation and Wiring
YSI 5X00
36
CAUTION: run high and low voltage cabling through separate bulkhead and
conduit
• where regular maintenance, including calibration can easily be done. Cali-
brations are performed using the 5X00 front panel.
Note:
- If a 5420 or 5421 sensor requires replacement, the entire probe/cable assembly
must to be replaced.
Locate sensor end of probe assembly:
• in a location that represents the entire system
• Insert the probe into sample. Move the probe in the sample to release any
air bubbles and to provide a fresh sample to the sensor cap. is movement
is only necessary when rst inserting the probe into the sample.
• Since the ProODO utilizes optical luminescent technology, continuous
sample movement or stirring is not required. (Please note however that ow
does increase response time of sensor.) e probe will t into a 300 mL BOD
bottle for taking initial and nal BOD readings. For best results in a BOD
bottle, a stirring device should be used to properly mix the sample and to
keep solids from settling at the bottom.
• Allow the temperature readings to stabilize and wait approximately 25-35
seconds for the DO readings to stabilize.
• in a location that is free from debris (e.g. algae); ideally at a 45˚ angle from
vertical position and directed with the ow.
• in a well mixed, free-owing area; midstream and mid-depth typically gives
representative readings; the probe assembly and cable are fully submersible
up to the cable connector.
• submersed at all times so liquid level uctuations do not expose the sensors
to the atmosphere.
• so the sensor/probe cabling is not routed near rotating machinery and/or
equipment involving electrical switching or regulation
• so that the sensor assembly and cabling are located away from sources of
electrical interference such as UV sterilizers, orescent lighting, ballasts,
pumps, any high voltage peripheral devices, etc.
CAUTION: run high and low voltage cabling through separate bulkhead and
conduit
5500
Installation and Wiring

Produktspecifikationer