Klein Tools CL2000 Bruksanvisning
Klein Tools
mätning
CL2000
Läs gratis den bruksanvisning för Klein Tools CL2000 (2 sidor) i kategorin mätning. Guiden har ansetts hjälpsam av 3 personer och har ett genomsnittsbetyg på 4.7 stjärnor baserat på 2 recensioner. Har du en fråga om Klein Tools CL2000 eller vill du ställa frågor till andra användare av produkten? Ställ en fråga
Sida 1/2

CUSTOMER SERVICE
KLEIN TOOLS, INC.
450 Bond Street
Lincolnshire, IL 60069
www.kleintools.com
139540T Rev. 11/13 B
ENGLISH
• TRUE RMS
• AUTO / MANUAL
RANGE
• NON CONTACT
VOLTAGE TESTER
• RELATIVE / ZERO
• BAR GRAPH
• MAX / MIN / HOLD
• 3-3/4 DIGIT
3999 COUNT LCD
• BACKLIGHT
• WORKLIGHT
V
AT
0 10 20 4030
Apo
600V
400A
CATIII
CL2000
ENGLISH
8. Non Contact Voltage (NCV):
>
25V AC
SYMBOLS USED ON LCD
~
AC Measurement DC Measurement
-
Negative DC Value
AT
Auto Range Active
O
.L.
Overload: Range Exceeded
Apo
Auto Power-Off Active
Low Battery
HOLD
Hold Active
MIN
Minimum Reading
M
AX
Maximum Reading
%
Duty Cycle Mode Frequency Mode
V
Voltage Measurement
A
Current in Amps
Ω
Resistance in Ohms Diode Test
F
Capacitance in Farads Continuity Test
Relative / Zero Mode
n
Nano 10
-9
µ
Micro 10
-6
m
Milli 10
-3
k
Kilo 10
3
M
Mega 10
6
ELECTRICAL SPECIFICATIONS
DC Voltage Measurement
Range Resolution Accuracy
400mV
˜
400V 0.1mV
˜
0.1V ± (0.5% + 4 digits)
1000V 1V ± (0.8% + 10 digits)
Overload Protection: 1000V
AC Voltage Measurement
Range Resolution Accuracy
400mV
˜
750V 0.1mV
˜
1V ± (2.0% + 5 digits)
Overload Protection: 750V RMS
Frequency Response: 40 to 400Hz
Minimum Voltage for Frequency Measurement: 200mV
Response: True RMS
Continuity Test
Overload Protection Open Circuit Voltage
600V RMS < 0.44V
Non Contact Voltage Detector
On Voltage
Appx. 25V AC
DC Current Measurement
Range Resolution Accuracy
400µA 0.1µA
± (1.2% + 3 digits)
2000µA 1µA
40A 0.01A ± (2.5% + 15 digits)
400A 0.1A ± (1.5% + 8 digits)
Overload Protection:
• Voltage: 600V RMS
• Current: 2000µA
AC Current Measurement
Range Resolution Accuracy
400µA 0.1µA ± (2.0% + 5 digits)
2000µA 1µA ± (1.5% + 5 digits)
40A 0.01A ± (2.9 % + 15 digits)
400A 0.1A ± (1.9% + 8 digits)
Overload Protection:
• Voltage: 600V RMS
• Current: 2000µA
Frequency: 45 to 400Hz
Minimum Current for Frequency Measurement: 400µA or 20A
Response: True RMS
Resistance Measurement
Range Resolution Accuracy
400
Ω
˜
4M 0.1 Ω Ω
˜
0.001MΩ
± (1.0% + 4 digits)
40M
Ω Ω0.01M
± (2.0% + 4 digits)
Overload Protection: 600V RMS
Capacitance Measurement
Range Resolution Accuracy
40nF
˜
4000µF 0.01nF
˜
1µF ± (3.5% + 6 digits)
Overload Protection: 600V RMS
Frequency Measurement
Range Resolution Accuracy
9.999Hz
˜
999.9kHz 0.001Hz
˜
0.1kHz ± (0.1% + 4 digits)
Overload Protection: 600V RMS
Duty Cycle Measurement
Range Resolution Accuracy
0.1
˜
99.9% 0.1% ± (0.2% per kHz + 0.1% + 5 digits)
Overload Protection: 600V RMS
Diode Test
Overload Protection Range Test Current Open Circuit Voltage
600V RMS 2.0V Appx. 0.25mA < 1.6V DC
600V
400V
CATIII
CL1000
PRESS
and
HOLD
PRESS
and
HOLD
WARRANTY
www.kleintools.com/warranty
CLEANING
Turn instrument off and disconnect test leads. Clean the instrument by using
a damp cloth. Do not use abrasive cleaners or solvents.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period of
time. Do not expose to high temperatures or humidity. After a period of storage
in extreme conditions exceeding the limits mentioned in the Specifications
section, allow the instrument to return to normal operating conditions before
using it.
DISPOSAL / RECYCLE
Caution: This symbol indicates that equipment and its accessories
shall be subject to a separate collection and correct disposal.
750V
1000V
400A
Instruction Manual
CL2000

ENGLISH
PRESS
V
AT
0 10 20 4030
Apo
PRESS
V
AT
0 10 20 4030
Apo
V
AT
0 10 20 4030
Apo
V
AT
0 10 20 4030
Apo
V
AT
0 10 20 4030
Apo
600V
400A
CATIII
CL2000
PRESS
FEATURE DETAILS
CL2000
Instruction Manual
GENERAL SPECIFICATIONS
The Klein Tools CL2000 is a True RMS, auto-ranging clamp meter.
It measures AC / DC voltage, AC / DC current, resistance, capacitance,
frequency, and duty cycle. It can also test non-contact voltage, diodes,
and continuity.
• 2000 metersOperating Altitude:
• 80% maxHumidity:
• Operating Temperature: 0°C / 32°F to 45°C / 113°F
• Storage Temperature: 0°C / 32°F to 60°C / 140°F
• Accuracy Temperature: 18°C / 64°F to 28°C / 82°F
• Temperature Coef cient: 0.1*(specified accuracy) / °C
• 8.875" x 3.375" x 1.5" (225 mm x 86 mm x 38 mm)Dimensions:
• 10.7 oz. (303 g)Weight:
• Accurate for one yearCalibration:
• CAT III 600VSafety Rating:
• ± (% of reading + # of least significant digits)Accuracy:
WARNINGS
To ensure safe operation and service of the tester, follow these instructions.
Failure to observe these warnings can result in severe injury or death.
• Before each use, verify meter operation by measuring
a known voltage or current.
• Never use the meter on a circuit with voltages that exceed the
category based rating of this meter.
• Do not use the meter during electrical storms, or in wet weather.
• Do not use the meter or test leads if they appear to be damaged.
• Ensure meter leads are fully seated, and keep fingers away from
the metal probe contacts when making measurements.
• Do not open the meter to replace batteries while the probes
are connected.
• Use caution when working with voltages above 60V DC,
or 25V AC RMS. Such voltages pose a shock hazard.
• To avoid false readings that can lead to electrical shock,
replace batteries if a low battery indicator appears.
• Unless measuring voltage or current, shut off and lock out
power before measuring resistance or capacitance.
• Always adhere to local and national safety codes. Use individual
protective equipment to prevent shock and arc blast injury where
hazardous live conductors are exposed.
SYMBOLS
~
AC Alternating Current Warning or Caution
DC Direct Current Dangerous levels
DC/AC Voltage or Current Double Insulated Class II
Ground
Safe for disconnect
from live conductors
AC Source
A. Use CAT III rated leads or higher.
Do not attempt to measure more than
1000V DC or 750V AC or 2000µA.
B. Keep hands below line when measuring high current levels.
C. Auto Power-Off (Apo)
• Device will power off after 30 minutes non-use.
• Turn the dial or press a button to wake.
• Disabled during Max / Min function.
• Holding Select button while turning on disables Auto Power-Off.
D. Select Functionality Button
• Switch between AC and DC.
• Switch between
and .
E. Hold / Backlight / Worklight
• Press to hold the current input on the display.
• Press again to return to live reading.
• Press for 2 seconds to enable disable lights. /
• Using lights drains the battery significantly.
F. Auto / Manual Range
• Press repeatedly to cycle through manual ranges.
• Press for 2 seconds to return to auto ranging mode.
•
AT is displayed on LCD only during auto ranging mode.
G. Max / Min Hold
• Press to enter Max / Min mode; the largest and smallest
values will be saved while in this mode.
• Press repeatedly to alternate between the maximum and minimum readings.
• Press for 2 seconds to return to live reading and clear
the stored maximum and minimum values.
H. I. Battery Replacement
• When indicator is displayed on the LCD, batteries must be replaced.
• Remove the back screw and replace 2 x AAA batteries.
J. Magnetic Back
• Attach instrument to metal for hands-free use.
K. Bar Graph
• The bar graph shows an approximate analog representation
of a measurement.
• The bar graph responds much faster than the digital display.
• The scale of the bar graph is zero to the maximum reading of
the selected range.
L. Relative Reading / Zero Mode
• Press for 2 seconds to enable/disable mode and store reading.
• Display will show the difference between the stored reading and
the live measurement.
FUNCTION INSTRUCTIONS
7
1
6
2
3
5
1. AC / DC Current (large):
<
400A
Features:
ZERO (DC)
REL (AC)
MAX/MINRANGEHOLD
• Center wire in guides for best accuracy.
• Opposing currents cancel. Use line-splitter (USA only) when necessary.
2. AC / DC Current (small):
<
2000µA
Features:
ZERO (DC)
REL (AC)
MAX/MINRANGEHOLD
• Select AC or DC current source.
•
Current above 2000µA will damage instrument.
3. AC / DC Voltage:
<
750V AC or 1000V DC
Features:
REL
MAX/MINRANGEHOLD
• Select AC or DC voltage source.
4. Frequency (Hz) / Duty Cycle (See Feature Details)
Features:
HOLD
• Select , , or setting.
• Frequencies greater than 1MHz will display “0.000Hz”.
5. Resistance:
<
40M
Ω
Features:
MAX/MINRANGEHOLD
• measure resistance on a live circuit.Do not
6. Capacitance:
<
4000µF
Features:
HOLD
• Safely discharge capacitor before measurement.
• Reading may take up to 60 seconds for large capacitors.
7. Diode / Continuity
Display shows:
• Forward voltage drop if forward biased.
• “O.L.” if reverse biased.
Diode Features:
• Display shows resistance.
• Buzzer sounds if less than 30
Ω.
Continuity Features:
HOLD MAX/MIN
TO SELECT
Diode
Continuity
PRESS
V
AT
0 10 20 4030
Apo
V
AT
0 10 20 4030
Apo
PRESS
Duty
Cycle
Hz
V
AT
0 10 20 4030
Apo
600V
400V
CATIII
B
A
I
8
J
4
600V
400V
CATIII
K
C
D
F
E
G
H
HOLD
MAX/MIN
L
1
Produktspecifikationer
Varumärke: | Klein Tools |
Kategori: | mätning |
Modell: | CL2000 |
Behöver du hjälp?
Om du behöver hjälp med Klein Tools CL2000 ställ en fråga nedan och andra användare kommer att svara dig
mätning Klein Tools Manualer
12 Juni 2025
12 Juni 2025
12 Juni 2025
12 Juni 2025
12 Juni 2025
12 Juni 2025
mätning Manualer
Nyaste mätning Manualer
14 Juni 2025
14 Juni 2025
13 Juni 2025
13 Juni 2025
9 Juni 2025
9 Juni 2025
9 Juni 2025
9 Juni 2025
9 Juni 2025
6 Juni 2025